Evaluating the Impact of Various Parameters on the Gamma Index Values of 2D Diode Array in IMRT Verification

(2018) Evaluating the Impact of Various Parameters on the Gamma Index Values of 2D Diode Array in IMRT Verification. J Med Signals Sens. pp. 31-38. ISSN 2228-7477 (Print) 2228-7477 (Linking)

Full text not available from this repository.

Abstract

Background: MapCHECK2 is a two-dimensional diode arrays planar dosimetry verification system. Dosimetric results are evaluated with gamma index. This study aims to provide comprehensive information on the impact of various factors on the gamma index values of MapCHECK2, which is mostly used for IMRT dose verification. Methods: Seven fields were planned for 6 and 18 MV photons. The azimuthal angle is defined as any rotation of collimators or the MapCHECK2 around the central axis, which was varied from 5 to -5 degrees . The gantry angle was changed from -8 to 8 degrees . Isodose sampling resolution was studied in the range of 0.5 to 4 mm. The effects of additional buildup on gamma index in three cases were also assessed. Gamma test acceptance criteria were 3/3 mm. Results: The change of azimuthal angle in 5 degrees interval reduced gamma index value by about 9. The results of putting buildups of various thicknesses on the MapCHECK2 surface showed that gamma index was generally improved in thicker buildup, especially for 18 MV. Changing the sampling resolution from 4 to 2 mm resulted in an increase in gamma index by about 3.7. The deviation of the gantry in 8 degrees intervals in either directions changed the gamma index only by about 1.6 for 6 MV and 2.1 for 18 MV. Conclusion: Among the studied parameters, the azimuthal angle is one of the most effective factors on gamma index value. The gantry angle deviation and sampling resolution are less effective on gamma index value reduction.

Item Type: Article
Keywords: Gamma index MapCHECK2 intensity modulated radiation therapy verification two-dimensional array
Divisions: Faculty of Medicine > Department of Basic Science > Department of Medical Physics
Faculty of Medicine > Departments of Clinical Sciences > Department of Radiooncology
Medical Image and Signal Processing Research Center
Page Range: pp. 31-38
Journal or Publication Title: J Med Signals Sens
Journal Index: Pubmed
Volume: 8
Number: 1
Identification Number: https://doi.org/10.4103/jmss.JMSS₁₅₁₇
ISSN: 2228-7477 (Print) 2228-7477 (Linking)
Depositing User: Zahra Otroj
URI: http://eprints.mui.ac.ir/id/eprint/7728