Error detection during VMAT delivery using EPID-based 3D transit dosimetry

(2018) Error detection during VMAT delivery using EPID-based 3D transit dosimetry. Physica Medica-European Journal of Medical Physics. pp. 137-145. ISSN 1120-1797

Full text not available from this repository.

Abstract

Purpose: To investigate the effectiveness of an EPID-based 3D transit dosimetry system in detecting deliberately introduced errors during VMAT delivery. Methods: An Alderson phantom was irradiated using four VMAT treatment plans (one prostate, two head-and-neck and one lung case) in which delivery, thickness and setup errors were introduced. EPID measurements were performed to reconstruct 3D dose distributions of "error" plans, which were compared with "no-error" plans using the mean gamma (gamma(mean)), near-maximum gamma (gamma 1) and the difference in isocenter dose (Delta D-isoc) as metrics. Results: Out of a total of 42 serious errors, the number of errors detected was 33 (79), and 27 out of 30 (90) if setup errors are not included. The system was able to pick up errors of 5 mm movement of a leaf bank, a wrong collimator rotation angle and a wrong photon beam energy. A change in phantom thickness of 1 cm was detected for all cases, while only for the head-and-neck plans a 2 cm horizontal and vertical shift of the phantom were alerted. A single leaf error of 5 mm could be detected for the lung plan only. Conclusion: Although performed for a limited number of cases and error types, this study shows that EPID-based 3D transit dosimetry is able to detect a number of serious errors in dose delivery, leaf bank position and patient thickness during VMAT delivery. Errors in patient setup and single leaf position can only be detected in specific cases.

Item Type: Article
Keywords: epid dosimetry vmat error detection quality assurance modulated arc therapy portal imaging device in-vivo dosimetry clinical-experience radiation-therapy quality-assurance prostate-cancer verification radiotherapy systems
Divisions: Faculty of Medicine > Department of Basic Science > Department of Medical Physics
Page Range: pp. 137-145
Journal or Publication Title: Physica Medica-European Journal of Medical Physics
Journal Index: ISI
Volume: 54
Identification Number: https://doi.org/10.1016/j.ejmp.2018.10.005
ISSN: 1120-1797
Depositing User: Zahra Otroj
URI: http://eprints.mui.ac.ir/id/eprint/9458

Actions (login required)

View Item View Item