Evaluating the effect of the vacuum bag on the dose distribution in radiation therapy

(2018) Evaluating the effect of the vacuum bag on the dose distribution in radiation therapy. Journal of Cancer Research and Therapeutics. pp. 1245-1250. ISSN 0973-1482

Full text not available from this repository.

Abstract

Introduction: Immobilization of patients in radiation therapy can be performed with a vacuum bag (VB). The aim of this study is to measure the effect of the VB in the surface and depth dose of patients in radiation therapy. Materials and Methods: The effect of the VB on the surface dose and depth dose is measured in clinical conditions. Various dosimeters were used in following measurements: parallel plate chamber for depth dose, farmer ionization chamber for various gantry angles, and Mapcheck2 dosimeter for various thicknesses of VB. The effect of VB cap, which may be placed in the beam path, is also measured using EDR2 film. The measurements were performed for 6 MV and 18 MV photons with an Oncor linac. Results: The increase of 30 and 25 in the surface dose with VB was observed for 6 MV and 18 MV, respectively. Though due to the use of VB, the reduction of the absorbed dose at a 5 cm depth is under 1 and can be ignored in MU calculation. For various thicknesses of VB, 8-14 cm, the attenuation of the primary beam were up to 2.5 for 6 MV and 1.2 for 18 MV photon. The presence of VB cap in the path of radiation reduced the depth dose up to 15 and 11 for 6 MV and 18 MV, respectively. Conclusion: The use of VB can increase the surface dose of the patient up to 30 and this fact should be considered in treatment planning. For some lateral fields the cap of the VB may interfere with radiation field. If the cap of VB is placed in the beam path, it can cause a cold spot in tumor. Discussion: The use of VB can increase the surface dose of the patient up to 30. For some lateral fields, the cap of the VB might interfere with the radiation field. If the cap of VB is placed in the beam path, it can cause a cold spot in the tumor volume.

Item Type: Article
Keywords: dosimetry immobilization radiation therapy vacuum bag quality-assurance patient immobilization prostate-cancer accuracy radiotherapy irradiation head alignment
Divisions: Faculty of Medicine > Department of Basic Science > Department of Medical Physics
Page Range: pp. 1245-1250
Journal or Publication Title: Journal of Cancer Research and Therapeutics
Journal Index: ISI
Volume: 14
Number: 6
Identification Number: Pmid 30488838 10.4103/0973-1482.188431
ISSN: 0973-1482
Depositing User: Zahra Otroj
URI: http://eprints.mui.ac.ir/id/eprint/9464

Actions (login required)

View Item View Item